Hipot Tester ST9201C

Ordernr. 60373
EAN 4260254474040
€1,750.00
Excl. 19% VAT, excl. Shipping Cost
Goods expected soon

Frequently Bought Together

  1. Hipot Tester ST9201C
  2. +
  3. Calibration HV-IS-OHM 2016
  4. +
  5. ST®EOL Fully automatic operating software
Total price: €0.00

Details
Technical Service
+49 421 277 9999

Highlights

  • Voltage: 5000V AC
  • 240×64 Dot-matrix graphic LCD
  • Adjustable ramp time (rise and fall)
  • Test steps programmable
  • GO-NOGO-Comparator
  • One view: nominal value & actual value
  • High safety: GR, GFI, fast discharge
  • RS-232C, Handler and SCAN Interface (optional)
  • Incl. HV accessories

Technical data

  • Memory: 50groups, 100 steps per group, totally 500 steps
  • Voltage rise-time: 0.1s - 999s
  • Voltage fall-time: 0.1s - 999s
  • Test time setting 0.3s - 999s
  •  and arc detection function
  • Body protection function
  • Built-in 8-channel matrix scanner for convenient use
  • Set voltage rising time, test time, and voltage dropping time
  • randomly for different load, DC withstanding voltage current
  • judging & waiting time
  • 100 test steps being stored per group, totally 50 groups, and the
  • total testing steps are limited at 500
  • Current base number correction function
  • Interfaces: USB, RS232, HANDLER, REMOTE I/O, SCAN (optional)
  • Power supply:  230V / 50 Hz
  • Dimensions (W/H/D) : 340/120/438.50 mm
  • Weight: 20kg
  • CE- conformity

Test methods

  • Withstanding voltage test
    - Output voltage
    - AC: 0.05kV — 5kV ± 1.0% of reading + 5 digit, (50Hz ,60Hz optional)
    - Voltage adjustment rate: ≤ 1.0% +10V (rated power)
    - Current test range
    - AC: 0.1mA - 20mA
     
  • Arc detection
    - Measurement range:     AC: 1mA - 15mA           

What's in the box:

  • ST90003R HV connection cable with test clip
  • ST90003B HV Earth connection cable with test clip
  • ST90002 HV Test probe set
  • Factory calibration certificate
  • Power cord
  • User manual

Technical specification

Model ST9201 ST9201S ST9201B ST9201C

Output voltage

AC Voltage 0.05KV - 5KV
Frequency 50Hz, 60Hz selectable
Output power 150VA (5kV/30mA) 100VA (5kV/20mA)
DC Voltage 0.05KV - 6KV 0.5KV - 6KV -
Output power 50VA (5kV/10mA) 25VA (5kV/5mA) -
Discharge function Automatically after end of test -
IR Voltage 50V-1000V -
Output power 10 VA (1000V/10mA) -
Discharge function Automatically after end of test -
8 Channel Matrix - Yes -

Voltage indication

Range Range of voltage 0.00kV – 6.00kV AC/DC
Accuracy ±(1.0% of reading +2V)
Range AC 0.001mA – 30mA 0.001mA – 20mA
DC 0.1µA – 10mA 0.1µA – 5mA -
Accuracy ±(1% of reading +10digits) (after correction)
ARC AC 1mA – 15mA 1mA – 15mA
DC 1mA – 10mA 1mA – 5mA
Insulation resistance indication
Resistance measuring range 0.01MΩ– 9.99GΩ, (Strom: 10nA – 10mA) -
Accuracy ±(5% of reading + 5digits) -

General specification

Voltage rise-time 0.1s - 999s
Voltage fall-time 0.1s - 999s
Voltage wait-time (DC) 0.1s - 99.9s (only for DC)
Test time setting 0.3s - 999s
More features Fast charge, body protection function, handler, RS-232, GPIB interface, USB host for memory sticks