Hipot Tester ST9201BX
€2,390.00*
Details
+49 421 277 9999
Highlights
- AC withstanding voltage and insulation tester
- Safety current limited
- 240×64 Dot-matrix graphic LCD display
- Fast discharge and arc detection function
- Body protection function
- Built-in 8-channel matrix scanner for convenient use
- Set voltage rising time, test time, and voltage dropping time
- randomly for different load, DC withstanding voltage current
- judging & waiting time
- 100 test steps being stored per group, totally 50 groups, and the
- total testing steps are limited at 500
- Current base number correction function
- RS-232C, Handler and SCAN Interface (optional)
Technical data
- 8-channel matrix scanner (not for 9201, 9201B, 9201C)
- Memory: 50groups, 100 steps per group, totally 500 steps
- Voltage rise-time: 0.1s - 999s
- Voltage fall-time: 0.1s - 999s
- Voltage wait-time: 0.1s - 99.9s (only for DC) (not 9201)
- Test time setting 0.3s - 999s
- Interfaces: USB, RS232, HANDLER, REMOTE I/O, SCAN (optional)
- Power supply: 230V / 50Hz
- Dimensions (W/H/D) : 340/120/438.50 mm
- Weight: 20kg
- CE - conformity
Test methods
Withstanding voltage test
- Output voltage
- AC: 0.05kV — 5kV ± 1.0% of reading+5 digit, (50Hz ,60Hz optional)
- DC: 0.05kV — 6kV ± 1.0% of reading+5 digit
- Voltage adjustment rate: ≤ 1.0% +10V (rated power)
- Current test range
- AC: 0.1mA - 3mA
- DC: 0.01mA - 5mA
- Test accuracy ± 1.0% of reading + 5 digit
- Discharge function: Discharge after test ends (DCW)
Insulation resistance test
- Output voltage:
- 0.05kV - 1kV ± (1.5% of reading + 5V) - Resistance test range:
- 0.1MΩ - 10GΩ, (Current range within 10nA - 10mA) - Resistance test accuracy:
- 500V-1000V 1MΩ - 1GΩ ±(5% of reading +5 digit) / 1GΩ - 10GΩ ±(10% of reading +5 digit)
- 50V-500V 0.1MΩ - 1GΩ ±(10% of reading +5 digit) - Discharge function: Discharge after test ends
Arc detection
- Measurement range:
- AC: 1mA - 3mA
- DC: 1mA - 5mA
What's in the box:
- ST90003R HV connection cable with test clip
- ST90003B HV Earth connection cable with test clip
- ST90002 HV Test probe set
- Factory calibration certificate
- Power cord
- User manual
HV Test: | AC/DC |
---|---|
Test current: | ≤3mA |
Test voltage: | >3kV - 6kV |
- ST9201BX Dimensions back (101.79 KB)
- ST9201BX Abmessungen Rückseite (101.79 KB)
- ST9201BX User Manual (1.97 MB)
- ST9201BX Bedienungsanleitung (2.06 MB)
Current indication
Technical specification |
|||||||||||
Model | ST9201SX | ST9201BX | ST9201CX | ||||||||
Output voltage |
|||||||||||
AC | Voltage | 0.05KV - 5KV | |||||||||
Frequency | 50Hz, 60Hz selectable | ||||||||||
Output power | 15VA (5kV/3mA) | 15VA (5kV/3mA) | |||||||||
DC | Voltage | 0.05KV - 6KV | 0.5KV - 6KV | - | |||||||
Output power | 50VA (5kV/10mA) | 25VA (5kV/5mA) | - | ||||||||
Discharge function | Automatically after end of test | - | |||||||||
IR | Voltage | 50V-1000V | - | ||||||||
Output power | 10 VA (1000V/10mA) | - | |||||||||
Discharge function | Automatically after end of test | - | |||||||||
8 Channel Matrix | Ja | - | |||||||||
Voltage indication |
|||||||||||
Range | Range of voltage | 0.00kV – 6.00kV AC/DC | |||||||||
Accuracy | ±(1.0% of reading +2V) | ||||||||||
Range | AC | 0.1mA – 3mA | 0.1mA – 3mA | ||||||||
DC | 0.01mA – 10mA | 0.01mA – 5mA | - | ||||||||
Accuracy | ±(1% of reading +10digits) (after correction) | ||||||||||
ARC | AC | 1mA – 3mA | 1mA – 3mA | ||||||||
DC | 1mA – 10mA | 1mA – 5mA | |||||||||
Insulation resistance indication | |||||||||||
Resistance measuring range | 0.01MΩ– 9.99GΩ, (Current: 10nA – 10mA) | - | |||||||||
Accuracy | ±(5% of reading + 5digits) | - | |||||||||
General specification |
|||||||||||
Voltage rise-time | 0.1s - 999s | ||||||||||
Voltage fall-time | 0.1s - 999s | ||||||||||
Voltage wait-time (DC) | 0.1s - 99.9s (only for DC) | ||||||||||
Test time setting | 0.3s - 999s | ||||||||||
More features | Fast charge, body protection function, handler, RS-232, GPIB interface, USB host for memory sticks |
Related products
- Efficient measurement of components and electrical devices
- Simple and intuitive operation
- Modular design: Integrate the test capabilities you need
- High flexibility in application
- Also compatible with third-party devices
- Standard-compliant, multi-vendor calibration based on ISO 17025
- Hipot and Insulationtester
- Traceable to DKD
- Cleaning and adjustment
- Repair service
- All-Round-Test-Cage to connect with hipot tester ST9200er series
- Dimension (W/H/D):
Outside: 450 / 515 / 495 mm
Inside: 360 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Devices: ST9201
- Voltage: 10.000 VAC
- Cross section: 4 mm²
- For up to 200°C
- Length: 1.50 m
- Robust 2-wire matrix
- Measuring points freely connectable
- Test connection via Harting plug
- Continuity tests
- High voltage tests
- Insulation resistance tests
- Surge voltage tests
- Devices: Hipot Tester ST9201, ST9201B/BX, ST9201C/CX, ST9201S/SX
- Robust transport "trolley"
- Aluminium frame lid can be opened
- Equipment compartment with punctual foam lining
- Clearly arranged accessory compartments
- Highest transport safety
- 2 middle butterfly locks at the front
- 2 locking hinges at the rear
- 2 centre folding handles
- Conversion kit of the high voltage tester ST9201
- Connection kit on the rear side
- Devices: ST9201
- Voltage: 10.000 VAC
- Cross section: 4 mm²
- For up to 200°C
- Length: 8.50 m
- Customized All-Round-Test-Cage to connect with hipot tester ST9200er Series
- Dimension (W/H/D):
Outside: 670 / 515 / 495 mm
Inside: 565 / 350 / 450 mm - Viewable transparent swinging hood
- Damping gas spring
- Safety limit switch
- CE conform with EN 50191
- Reliable HV Probe
- Highly flexible HV connecting cable
- Operating voltage 10 kVDC / 7 kVAC
- Hard-wearing tungsten test prod
- CE conformity, according EN 50 191
- Devices: series ST9200, ST9300
- Cable length: 1.5m
- Standard: EN 50191
- Lights: Red/Green
- Function: rd HV ON/gr HV OFF
- 19" mounting frame
- Color: RAL light grey
- Height unit: 4HE
- Devices: ST9201 series